This is a high speed digital servo-controlled, volume constant system with a monitor manifold parameters, suitable for adult & paediatric and comes with a compliance and leakage compensation system

Category:

This is a high speed digital servo-controlled, volume constant system with a monitor manifold parameters, suitable for adult & paediatric and comes with a compliance and leakage compensation system

Features

  • TFT screen display
  • Self testing, troubleshooting and warning system
  • High quality vaporizers: Low, temperature and pressure compensation
  • Low resistance, compact absorber system
  • Special designed bellows joint for easy installation
  • Precision processing
  • High reliability and durability

Main Unit

  • Mode: Semi-open, semi-closed or closed system
  • Gas supply: O2/N2O, Pressure: 0.3 to 0.5 MPa
  • Flowmeter: O2/N2O four-tube flowmeter, 0 to 10 L/min
  • (With inter-lock protection devicesensuring the oxygen concentration > 25 %)
    O2flush: 35 to 75 L/min
  • O2deficiency: Audible alarm for at least 7 seconds when O2supply down

MV220 Ventilator

  • Patient type: Adult, Paediatric
  • Driven mode: Pneumatically driven electronically controlled
  • Display mode: 5.7″TFT screen display
  • Ventilation mode: Volume, Manual
  • Tidal volume: 0 to 1500 ml
  • Breath rate: 4 to 99 bpm
  • I: E ratio: 1: 0.5 to 1: 6
  • Pause time: 0 to 50% ( increments of 5% )
  • PEEP: 0.3 to 2 kPa (optional)
  • Alarm silence: ≤120 seconds
  • Monitoring: Tidal volume, MV, Breath rate, Airway pressure waveform, Flow waveform, Airway pressure(peak, plat, average), Dynamic compliance, O2concentration, PEEP (optional)
  • Alarm: High/Low airway pressure, Vt high, Airway failure, MV high/low, Apnea, AC powerfailure
    Safety valve: ≤6 kPa

VP200 Vaporizer

  • Capacity: 120 ml (Max. )
  • Dry wick Volume: 20 ml
  • Flow range: 0.2 to 15 L/min
  • Concentration: 0 to 5.0%
  • Work Temp.: 15to 35
  • Storage Temp.: – 40to 70
  • Dimension: 155 mm*94 mm*189 mm(D*W*H)
  • Weight: 5 kg

Operation Environment

  • Power supply: 110to 240 V; 50to 60 Hz
  • Power consumption: ≤50 VA
  • Operation Temp.: 10to40
  • Storage Temp.: – 10to 55
  • Humidity: ≤ 93%
  • Dimension and Weight
  • Dimension: 620 mm*910 mm*1387 mm(D*W*H)
  • Weight: 97 kg